All qualifications and part qualifications registered on the National Qualifications Framework are public property. Thus the only payment that can be made for them is for service and reproduction. It is illegal to sell this material for profit. If the material is reproduced or quoted, the South African Qualifications Authority (SAQA) should be acknowledged as the source. |
SOUTH AFRICAN QUALIFICATIONS AUTHORITY |
REGISTERED UNIT STANDARD THAT HAS PASSED THE END DATE: |
Calibrate oscilloscopes |
SAQA US ID | UNIT STANDARD TITLE | |||
119248 | Calibrate oscilloscopes | |||
ORIGINATOR | ||||
SGB Engineering | ||||
PRIMARY OR DELEGATED QUALITY ASSURANCE FUNCTIONARY | ||||
- | ||||
FIELD | SUBFIELD | |||
Field 06 - Manufacturing, Engineering and Technology | Engineering and Related Design | |||
ABET BAND | UNIT STANDARD TYPE | PRE-2009 NQF LEVEL | NQF LEVEL | CREDITS |
Undefined | Regular | Level 5 | Level TBA: Pre-2009 was L5 | 18 |
REGISTRATION STATUS | REGISTRATION START DATE | REGISTRATION END DATE | SAQA DECISION NUMBER | |
Passed the End Date - Status was "Reregistered" |
2018-07-01 | 2023-06-30 | SAQA 06120/18 | |
LAST DATE FOR ENROLMENT | LAST DATE FOR ACHIEVEMENT | |||
2026-06-30 | 2029-06-30 |
In all of the tables in this document, both the pre-2009 NQF Level and the NQF Level is shown. In the text (purpose statements, qualification rules, etc), any references to NQF Levels are to the pre-2009 levels unless specifically stated otherwise. |
This unit standard does not replace any other unit standard and is not replaced by any other unit standard. |
PURPOSE OF THE UNIT STANDARD |
This unit standard will be useful to people who are required to perform measurements for the purpose of calibrating oscilloscopes. Persons credited with this unit standard will have the generic skills to calibrate general purpose oscilloscopes. This includes the calibration of all oscilloscopes functions by performing measurements and maintaining the necessary standards and transferring accuracy traceability from National Measurement Standards for DC and AC Voltage, Frequency, Rise time, to a unit under test.
|
LEARNING ASSUMED TO BE IN PLACE AND RECOGNITION OF PRIOR LEARNING |
Competence in the following:
|
UNIT STANDARD RANGE |
N/A |
Specific Outcomes and Assessment Criteria: |
SPECIFIC OUTCOME 1 |
Maintain oscilloscope calibration reference standards. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
Evidence of the source of traceability selected, has international equivalence. |
ASSESSMENT CRITERION 2 |
Calibration intervals are determined from analysed data of reference standards' records. |
ASSESSMENT CRITERION 3 |
Reference standards are maintained in a condition fit for purpose. |
ASSESSMENT CRITERION 4 |
Reference standards are verified to be fit for intended purpose, data recorded and corrective action implemented for failures. |
ASSESSMENT CRITERION 5 |
Maintenance records for reference standards have been generated and kept up to date. |
ASSESSMENT CRITERION RANGE |
"Maintenance records" refers to calibration certificate; record cards; verification data; control charts; etc |
ASSESSMENT CRITERION 6 |
Laboratory environmental conditions are within required limits and records are kept. |
ASSESSMENT CRITERION 7 |
Reference standards are maintained to within an uncertainty in support of the required measurement capability. |
ASSESSMENT CRITERION RANGE |
"Reference standards" refers to national measurement standards, laboratory standards, working standards and shop-floor standards. |
SPECIFIC OUTCOME 2 |
Calibrate oscilloscope display. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
Explain the principles of oscilloscope display as it is affected by beam focus, beam astigmatism, pin-cushion distortion, vertical and horizontal non-linearity. |
ASSESSMENT CRITERION 2 |
Calibrate an oscilloscope to optimize the display characteristics. |
ASSESSMENT CRITERION 3 |
Perform measurements according to a given procedure. |
SPECIFIC OUTCOME 3 |
Calibrate oscilloscope vertical systems. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
Explain the principles of operation of an oscilloscope vertical system. |
ASSESSMENT CRITERION 2 |
Select the appropriate reference standards and calibrate the oscilloscope vertical gain, bandwidth, rise time and input impedance according to given procedure. |
ASSESSMENT CRITERION 3 |
Record the measurement results. |
ASSESSMENT CRITERION 4 |
Estimate the uncertainties of measurement for the measurements performed. |
SPECIFIC OUTCOME 4 |
Calibrate oscilloscope horizontal systems. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
Explain the principles of operation of an oscilloscope horizontal system. |
ASSESSMENT CRITERION 2 |
Select the appropriate reference standards and calibrate the oscilloscope horizontal gain, bandwidth, according to given procedure. |
ASSESSMENT CRITERION 3 |
Record the measurement results. |
ASSESSMENT CRITERION 4 |
Estimate the uncertainties of measurement for the measurements performed. |
SPECIFIC OUTCOME 5 |
Calibrate oscilloscope trigger systems. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
Explain the principles of operation of an oscilloscope trigger system. |
ASSESSMENT CRITERION 2 |
Select the appropriate reference standards and calibrate the oscilloscope trigger sensitivity, polarity, according to given procedure. |
ASSESSMENT CRITERION 3 |
Record the measurement results. |
ASSESSMENT CRITERION 4 |
Estimate the uncertainties of measurement for the measurements performed. |
SPECIFIC OUTCOME 6 |
Calibrate oscilloscope Z-axis system. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
Explain the principles of operation of an oscilloscope Z-axis system. |
ASSESSMENT CRITERION 2 |
Select the appropriate reference standards and calibrate the oscilloscope Z-axis bandwidth and gain according to given procedure. |
ASSESSMENT CRITERION 3 |
Record the measurement results. |
ASSESSMENT CRITERION 4 |
Estimate the uncertainties of measurement for the measurements performed. |
SPECIFIC OUTCOME 7 |
Calibrate oscilloscope X-Y system. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
Explain the principles of operation of an oscilloscope X-Y system. |
ASSESSMENT CRITERION 2 |
Select the appropriate reference standards and calibrate the oscilloscope X-Y system bandwidth and phase difference according to given procedure. |
ASSESSMENT CRITERION 3 |
Record the measurement results. |
ASSESSMENT CRITERION 4 |
Estimate the uncertainties of measurement for the measurements performed. |
SPECIFIC OUTCOME 8 |
Record, evaluate and report results. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
Record raw data as per procedure. |
ASSESSMENT CRITERION 2 |
Evaluate raw data for conformance to required accuracy specifications. |
ASSESSMENT CRITERION 3 |
Make adjustments, if necessary. |
ASSESSMENT CRITERION 4 |
Generate an appropriately detailed measurement report/certificate in accordance with ISO/IEC 17025. |
UNIT STANDARD ACCREDITATION AND MODERATION OPTIONS |
Anyone assessing a learner against this unit standard must be a registered SANAS Technical Assessor and registered as an assessor with the relevant ETQA.
Any institution offering learning that will enable achievement of this unit standard must be accredited as a provider through the relevant ETQA by SAQA. Moderation of assessment will be overseen by the relevant ETQA according to the moderation guidelines and the agreed ETQA procedures. |
UNIT STANDARD ESSENTIAL EMBEDDED KNOWLEDGE |
Essential embedded knowledge will be assessed through assessment of the specific outcomes in terms of the stipulated assessment criteria. Candidates are unlikely to achieve all the specific outcomes, to the standards described in the assessment criteria, without knowledge of the listed embedded knowledge. This means that for the most part, the possession or lack of the knowledge can be directly inferred from the quality of the candidate's performance. Where direct assessment of knowledge is required, assessment criteria have been included in the body of the unit standard.
The following embedded knowledge is addressed in an integrated way in the unit standard: |
UNIT STANDARD DEVELOPMENTAL OUTCOME |
N/A |
UNIT STANDARD LINKAGES |
N/A |
Critical Cross-field Outcomes (CCFO): |
UNIT STANDARD CCFO IDENTIFYING |
Identify and solve problems and make decisions using critical and creative thinking.
Note: The ability of the candidate to identify incorrect conditions with reference standards, incorrect measurements taken for a specific parameter, measurements outside specifications, assess and take appropriate action - Specific Outcome |
UNIT STANDARD CCFO WORKING |
Work effectively with others as members of a team, group, organisation or community.
Note: The ability and willingness of the candidate to accept and interpret work instructions and calibration/measurement procedures from customer's correctly and to request assistance in an appropriate manner as a team member Specific outcomes |
UNIT STANDARD CCFO ORGANISING |
Organise and manage themselves and their activities responsibly and effectively.
Note: The preparation for calibration of oscilloscopes, the maintenance of reference standards, the candidate must indicate what tools, equipment, set-up and calibration procedures are required and communicates to fellow workers and laboratory manager his/her intentions Specific outcomes |
UNIT STANDARD CCFO COLLECTING |
Collect, analyse, organise and critically evaluate information.
Note: The ability of the candidate to reconcile the measured data and information from the visual examination and constantly evaluate the changing situation Specific outcome |
UNIT STANDARD CCFO COMMUNICATING |
Communicate effectively, using visual, mathematical and / or language skills in the modes of oral and / or written presentations.
Note: The appropriate communication with customers, laboratory manager and personnel with regard to maintenance of Reference Standards, calibration and/or testing of Oscilloscopes and the measurement results achieved, will indicate his/her proficiency in effective communication Specific outcomes |
UNIT STANDARD CCFO SCIENCE |
Use science and technology effectively and critically showing responsibility towards the environment and health of others.
Note: The ability of the candidate to effectively use the knowledge gained to inform and protect the broader environment and health of others Specific outcomes |
UNIT STANDARD CCFO DEMONSTRATING |
Demonstrate an understanding of the world as a set of related systems by recognising that problem-solving contexts do not exist in isolation.
Note: The ability of the candidate to understand the inter relationship between electrical measurements, maintenance and calibration of standards, reporting of results and their effect on global activities Specific outcomes |
UNIT STANDARD ASSESSOR CRITERIA |
N/A |
REREGISTRATION HISTORY |
As per the SAQA Board decision/s at that time, this unit standard was Reregistered in 2012; 2015. |
UNIT STANDARD NOTES |
Assessors should keep the following principles in mind when designing and conducting assessments against this unit standard:
Unit standard justification 1. Level Attribute - Level Applied Competence - 5 Autonomy of Learning - 5 Actual Level Assigned: 5 2. Credits Total hours required by the learner to achieve the required outcome: Activity-Hours Classroom Teaching-40 On-The-Job Training-100 Mentoring required-40 Other (Specify)-- Total-180 Credits Assigned: 18 |
QUALIFICATIONS UTILISING THIS UNIT STANDARD: |
ID | QUALIFICATION TITLE | PRE-2009 NQF LEVEL | NQF LEVEL | STATUS | END DATE | PRIMARY OR DELEGATED QA FUNCTIONARY | |
Elective | 57228 | National Certificate: Metrology | Level 5 | Level TBA: Pre-2009 was L5 | Passed the End Date - Status was "Reregistered" |
2023-06-30 | EWSETA |
PROVIDERS CURRENTLY ACCREDITED TO OFFER THIS UNIT STANDARD: |
This information shows the current accreditations (i.e. those not past their accreditation end dates), and is the most complete record available to SAQA as of today. Some Primary or Delegated Quality Assurance Functionaries have a lag in their recording systems for provider accreditation, in turn leading to a lag in notifying SAQA of all the providers that they have accredited to offer qualifications and unit standards, as well as any extensions to accreditation end dates. The relevant Primary or Delegated Quality Assurance Functionary should be notified if a record appears to be missing from here. |
NONE |
All qualifications and part qualifications registered on the National Qualifications Framework are public property. Thus the only payment that can be made for them is for service and reproduction. It is illegal to sell this material for profit. If the material is reproduced or quoted, the South African Qualifications Authority (SAQA) should be acknowledged as the source. |