All qualifications and part qualifications registered on the National Qualifications Framework are public property. Thus the only payment that can be made for them is for service and reproduction. It is illegal to sell this material for profit. If the material is reproduced or quoted, the South African Qualifications Authority (SAQA) should be acknowledged as the source. |
SOUTH AFRICAN QUALIFICATIONS AUTHORITY |
REGISTERED UNIT STANDARD THAT HAS PASSED THE END DATE: |
Calibrate isothermal media by temperature simulation and temperature mapping at temperatures not exceeding 250ºc |
SAQA US ID | UNIT STANDARD TITLE | |||
230192 | Calibrate isothermal media by temperature simulation and temperature mapping at temperatures not exceeding 250ºc | |||
ORIGINATOR | ||||
SGB Engineering | ||||
PRIMARY OR DELEGATED QUALITY ASSURANCE FUNCTIONARY | ||||
- | ||||
FIELD | SUBFIELD | |||
Field 06 - Manufacturing, Engineering and Technology | Engineering and Related Design | |||
ABET BAND | UNIT STANDARD TYPE | PRE-2009 NQF LEVEL | NQF LEVEL | CREDITS |
Undefined | Regular | Level 5 | Level TBA: Pre-2009 was L5 | 10 |
REGISTRATION STATUS | REGISTRATION START DATE | REGISTRATION END DATE | SAQA DECISION NUMBER | |
Passed the End Date - Status was "Reregistered" |
2018-07-01 | 2023-06-30 | SAQA 06120/18 | |
LAST DATE FOR ENROLMENT | LAST DATE FOR ACHIEVEMENT | |||
2026-06-30 | 2029-06-30 |
In all of the tables in this document, both the pre-2009 NQF Level and the NQF Level is shown. In the text (purpose statements, qualification rules, etc), any references to NQF Levels are to the pre-2009 levels unless specifically stated otherwise. |
This unit standard does not replace any other unit standard and is not replaced by any other unit standard. |
PURPOSE OF THE UNIT STANDARD |
Since Metrology covers a wide spectrum of measurement disciplines, this unit standard should be seen as covering only one of the disciplines. Although many metrology processes are similar, there are distinct competency differences between various technically specialized parameters within each discipline.
This unit standard will be useful to metrologists who are required to perform measurements and calibrations within the TEMPERATURE discipline in the following parameters: > Temperature Simulation and Mapping of Isothermal media. (Temperatures will typically range from sub-zero to 250ºC) Persons credited with this unit standard are able to calibrate devices and artefacts for metrology. This includes all aspects related to the work of performing measurements and maintaining the necessary standards and transferring accuracy traceability from National Measurement Standards to a unit under test. The Metrology process, referred to above, is generically the same for all disciplines within metrology fields, but the embedded knowledge is uniquely different. The process is to: When Isothermal Media are to be calibrated, the generic Metrology process mentioned above, will focus specifically on temperature simulation and mapping which includes: Therefore, the purpose of this unit standard is to calibrate Isothermal media and perform temperature mapping for metrology, including all aspects of using temperature calibrators to simulate stable temperatures; calibrating temperature indicating, controlling and recording devices at specific temperature points for transferring accuracy traceability. Learners will be able to: |
LEARNING ASSUMED TO BE IN PLACE AND RECOGNITION OF PRIOR LEARNING |
It is assumed that learners are already competent in:
|
UNIT STANDARD RANGE |
N/A |
Specific Outcomes and Assessment Criteria: |
SPECIFIC OUTCOME 1 |
Simulate known temperatures electrically. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
The fundamentals of Electrically simulated Temperatures, are explained with examples. |
ASSESSMENT CRITERION 2 |
The simulation source is selected relative to a source of traceability with national equivalence. |
ASSESSMENT CRITERION 3 |
Calibration intervals are established from analysed data of reference standards' records. |
ASSESSMENT CRITERION 4 |
The Simulation source is verified fit for intended purpose, and maintenance records are updated. |
ASSESSMENT CRITERION RANGE |
"Maintenance records" refers to calibration certificate; record cards; verification data; control charts; etc. |
ASSESSMENT CRITERION 5 |
Voltages and resistances are generated to simulate a given temperature for a given Resistive Temperature Device/Thermocouple and loggers/recorders. |
ASSESSMENT CRITERION 6 |
The environmental conditions, where calibrations are performed, are recorded in accordance with procedures. |
ASSESSMENT CRITERION 7 |
The simulation source is verified against reference standards before and after on-site calibration work is performed, to prove required accuracy. |
ASSESSMENT CRITERION 8 |
Simulated temperatures generated, are confirmed by measurement. |
SPECIFIC OUTCOME 2 |
Verify Iso-thermal device temperatures. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
The thermocouples/RTD's are verified against an ice point before and after on-site calibration work is performed, to prove required accuracy. |
ASSESSMENT CRITERION 2 |
Measurements required are identified in accordance with isothermal media devices. |
ASSESSMENT CRITERION 3 |
Temperatures are measured in a given isothermal device over sufficient time, to identify cyclic temperature behaviour, and at suitable positions to map the temperature profile of the device. |
ASSESSMENT CRITERION NOTES |
On Sterilising Autoclaves, it is necessary to perform limited Time Interval and Pressure calibration measurements in addition to the temperature mapping measurements described above. |
ASSESSMENT CRITERION RANGE |
ASSESSMENT CRITERION 4 |
The measurement method is selected and followed in accordance with procedures. |
ASSESSMENT CRITERION 5 |
Calibration points are selected to satisfy calibration requirements. |
ASSESSMENT CRITERION 6 |
Measurements are performed within the required precision. |
ASSESSMENT CRITERION 7 |
Applicable safety requirements are adhered to in accordance with current legislation. |
SPECIFIC OUTCOME 3 |
Record, evaluate and report results. |
ASSESSMENT CRITERIA |
ASSESSMENT CRITERION 1 |
The raw measurement data is recorded in accordance with requirements. |
ASSESSMENT CRITERION 2 |
The uncertainty of raw measurement data is estimated in accordance with applicable standards. |
ASSESSMENT CRITERION 3 |
The raw data is evaluated for conformance to given accuracy specifications. |
ASSESSMENT CRITERION 4 |
An appropriately detailed measurement report/certificate is generated for given measurement results in accordance with applicable standards. |
UNIT STANDARD ACCREDITATION AND MODERATION OPTIONS |
UNIT STANDARD ESSENTIAL EMBEDDED KNOWLEDGE |
Essential embedded knowledge will be assessed through assessment of the specific outcomes in terms of the stipulated assessment criteria. Candidates are unlikely to achieve all the specific outcomes, to the standards described in the assessment criteria, without knowledge of the listed embedded knowledge. This means that for the most part, the candidate's performance. Where direct assessment of knowledge is required, assessment criteria have been included in the body of the unit standard.
The following embedded knowledge is addressed in an integrated way in the unit standard: Laws of thermodynamics. Ice point, laws of intermediate temperatures and intermediate metals, use of Reference Junctions/loggers. Specification of isothermal devices, Design principle, accuracy requirements, Temperature verification, Microvolt/Resistance measurement by mapping. > Calibration of Sterilising Autoclaves. Pressure calibration: calibration of pressure gauges using reference pressure gauge and manifold. Calibration of timers: Time interval calibration using reference stopwatch/electronic counter. > Calibration of Drying Ovens, Incubators, Sterilisers and Environmental Chambers. Specification of temperature sensors to achieve desired accuracy. > Calibration of Fridges and Freezers. Sub-zero temperature measurement principles and techniques. > Calibration of Thermal Block Digestors. Ramping temperature measurement in test samples. |
UNIT STANDARD DEVELOPMENTAL OUTCOME |
N/A |
UNIT STANDARD LINKAGES |
N/A |
Critical Cross-field Outcomes (CCFO): |
UNIT STANDARD CCFO IDENTIFYING |
Identify and solve problems and make decisions using critical and creative thinking.
Specific Outcome 1, 2 and 3 |
UNIT STANDARD CCFO WORKING |
Work effectively with others as members of a team, group, organisation or community.
Specific Outcome 3 |
UNIT STANDARD CCFO ORGANISING |
Organise and manage themselves and their activities responsibly and effectively.
Specific Outcome 1, 2 and 3 |
UNIT STANDARD CCFO COLLECTING |
Collect, analyse, organise and critically evaluate information.
Specific Outcome 1, 2 and 3 |
UNIT STANDARD CCFO COMMUNICATING |
Communicate effectively, using visual, mathematical and/or language skills in the modes of oral and/or written presentations.
|
UNIT STANDARD CCFO SCIENCE |
Use science and technology effectively and critically showing responsibility towards the environment and health of others.
Specific Outcome 1, 2 and 3 |
UNIT STANDARD CCFO DEMONSTRATING |
Demonstrate an understanding of the world as a set of related systems by recognising that problem-solving contexts do not exist in isolation.
Specific Outcome 1, 2 and 3 |
UNIT STANDARD ASSESSOR CRITERIA |
N/A |
REREGISTRATION HISTORY |
As per the SAQA Board decision/s at that time, this unit standard was Reregistered in 2012; 2015. |
UNIT STANDARD NOTES |
Notes to assessors:
Assessors should keep the following principles in mind when designing and conducting assessments against this unit standard: Unit standard justification: 1. Level Actual Level Assigned: 5 2. Credits Total hours required by the learner to achieve the required outcome: |
QUALIFICATIONS UTILISING THIS UNIT STANDARD: |
ID | QUALIFICATION TITLE | PRE-2009 NQF LEVEL | NQF LEVEL | STATUS | END DATE | PRIMARY OR DELEGATED QA FUNCTIONARY | |
Elective | 57228 | National Certificate: Metrology | Level 5 | Level TBA: Pre-2009 was L5 | Passed the End Date - Status was "Reregistered" |
2023-06-30 | EWSETA |
PROVIDERS CURRENTLY ACCREDITED TO OFFER THIS UNIT STANDARD: |
This information shows the current accreditations (i.e. those not past their accreditation end dates), and is the most complete record available to SAQA as of today. Some Primary or Delegated Quality Assurance Functionaries have a lag in their recording systems for provider accreditation, in turn leading to a lag in notifying SAQA of all the providers that they have accredited to offer qualifications and unit standards, as well as any extensions to accreditation end dates. The relevant Primary or Delegated Quality Assurance Functionary should be notified if a record appears to be missing from here. |
NONE |
All qualifications and part qualifications registered on the National Qualifications Framework are public property. Thus the only payment that can be made for them is for service and reproduction. It is illegal to sell this material for profit. If the material is reproduced or quoted, the South African Qualifications Authority (SAQA) should be acknowledged as the source. |